  • Page 1 Keysight M8290A Optical Modulation Analyzer and High-Speed Digitizer Test Solution Getting Started Guide...
  • Page 2: Safety Notices

    DFARS and are set forth specifically in Technology Licenses not correctly performed or adhered writing elsewhere in the EULA. Keysight to, could result in damage to the The hardware and/or software described in shall be under no obligation to update, product or loss of important data.
  • Page 3: Safety Summary

    M9505A AXIe chassis, must be followed. See: http://www.keysight.com/find/M9505A Keysight Technologies Inc. assumes no liability for the customer's failure to comply with these requirements. Before operation, review the instrument and manual for safety markings and instructions. You must follow these to ensure safe operation and to maintain the instrument in safe condition.
  • Page 4 Line Power Requirements The Keysight M829xA modules operate when installed in a Keysight AXIe mainframe. Do Not Operate in an Do not operate the instrument in the presence of flammable gases or fumes. Explosive Atmosphere Do Not Remove the Operating personnel must not remove instrument covers. Component replacement and internal adjustments must be made only by qualified personnel.
  • Page 5: General Safety Considerations

    The laser radiation symbol. This warning symbol is marked on products which have a laser output. The electrostatic discharge symbol. This warning symbol is marked on products which have components that can be damaged by an electrostatic discharge. Keysight M8290A Getting Started Guide...
  • Page 6 CE compliance marking to the EU Safety and EMC Directives. ISM GRP-1A classification according to the international EMC standard. ICES/NMB-001 compliance marking to the Canadian EMC standard. This symbol on all primary and secondary packaging indicates compliance to China standard GB 18455-2001. Keysight M8290A Getting Started Guide...
  • Page 7: Compliance And Environmental Information

    Product Category: With reference to the equipment types in WEEE Directive Annex I, this product is classed as a “Monitoring and Control instrumentation” product. Do not dispose in domestic household waste. To return unwanted products, contact your local Keysight office, or see http://about.keysight.com/en/companyinfo/environment/takeback.shtml for more information.
  • Page 8 Keysight M8290A Getting Started Guide...
  • Page 9: Table Of Contents

    3 Using the Instrument How to use the Instrument Adding the M829xA to the Connection Expert Setting up the M829xA in the Keysight OMA Software 4 AXle Chassis ESM Front Panel Connector Setting the Startup Delay for the Embedded Controller...
  • Page 10 Initial Safety Information Laser Safety Labels ESD Protection Power and Ventilation Requirements Thermal Protection Battery Cleaning Recommendation 6 Specifications M8292A Modular Optical Modulation Analyzer Specifications M8296A Electrical High-Speed Digitizer Specifications General Characteristics Declaration of Conformity Keysight M8290A Getting Started Guide...
  • Page 11: Introduction Getting Started Additional Documents

    / 12 Additional Documents Introduction The M8290A rack-mountable modular coherent test system for the AXIe platform is designed to address the 400G speed class in a significantly narrower form factor and a more attractive price point than today’s oscilloscope-based solutions for this speed class.
  • Page 12 The M8290A Coherent Test Solution comprises an M9537A AXIe embedded controller. This controller can control one or multiple M829xA and consumes one module slot of the AXIe chassis. The M8290A Coherent Test Solution comes with all required software pre-installed and configured at the factory.
  • Page 13: Pre-Requisites

    Installing the Keysight OMA Software / 27 Post Installation Steps The M8290A coherent test solution uses a layered software architecture, which is illustrated below, that requires multiple software packages to be installed. This chapter explains how to install and configure these packages.
  • Page 14 • Windows 7 (32 bit or 64 bit) • Ensure that you have Keysight IO Libraries Suite Version 17.3 or higher installed on your system. The Keysight IO Libraries Suite can be found at http://www.keysight.com/find/iosuite. Keysight M8290A Getting Started Guide...
  • Page 15 Software Installation Even if a non-Keysight I/O library is already installed on your PC, it is still NOTE necessary to install the Keysight I/O library. The Keysight I/O library will install as “secondary” I/O library in this case. This use case is fully supported.
  • Page 16 Software Installation Click on the Open button to open the zip file. Unzip the file and double-click on the executable file to run the PAS Package Manager. Keysight M8290A Getting Started Guide...
  • Page 17 Software Installation Start the PAS Package Manager as administrator and select the OMA tab to see all possible OMA packages. Keysight M8290A Getting Started Guide...
  • Page 18: M8290A Software Installation

    Software Installation M8290A Software Installation Follow the steps below to install the Keysight M8290A software on your system: Start the PAS Package Manager as administrator, if it is not already open and select the OMA tab to see all possible OMA packages.
  • Page 19 Software Installation The Keysight M8290A Setup will prepare the InstallShield Wizard for the installation process. The following screen will appear. Keysight M8290A Getting Started Guide...
  • Page 20 Software Installation Click Next. The following screen will appear: Keysight M8290A Getting Started Guide...
  • Page 21 Software Installation Click Next to proceed to the license agreements. Accept the terms of ‘Keysight Software End-User License Agreement’ and click Next. The following screen will appear: Select Yes if you want to read the post-installation instructions and click Next. The following screen will appear:...
  • Page 22 10 Select a setup type of either Complete or Custom. If you select Complete, the M8290A documentation will be installed along with the M8290A software and a shortcut to the PAS Package Manager will be added to the Start menu. If you select Custom instead, you can specify which of the optional features to install.
  • Page 23 Software Installation 12 Click Install to begin the software installation. The Setup Wizard will now install the M8290A software. Keysight M8290A Getting Started Guide...
  • Page 24 Software Installation 13 The following screen will appear once the Keysight M8290A software is successfully installed on your system. 14 Click Finish to restart your system. This completes the Keysight M8290A software installation. Keysight M8290A Getting Started Guide...
  • Page 25 The PAS Package Manager will install additional drivers and software required to install the OMA software. During the OMA installation, select M8290A Hardware as setup type. You may postpone all system restarts until the installation has been completed.
  • Page 26 Software Installation Keysight M8290A Getting Started Guide...
  • Page 27 Start > Control Panel > Device Manager, or right-click Computer > Manage > Device Manager. The instruments should be visible in the device tree as Keysight Technologies Modular Devices > M8296A or M8292A respectively. Check if all installed M829xA modules are visible in the Connection Expert.
  • Page 29: How To Use The Instrument

    Setting up the M829xA in the Keysight OMA Software on page 32. To show in the Keysight OMA Software, the module must have been NOTE added to the Connection Expert as a LAN instrument. Socket connections are not supported.
  • Page 30 Using the Instrument Adding the M829xA to the Connection Expert A standard M8290A system is pre-configured at the factory. This section shows how to modify the factory configuration or to configure a stand-alone module. Open the Connection Expert by clicking its icon ( ) in the system tray and selecting the corresponding item from the context menu.
  • Page 31 “My Instruments” section. If so, the instrument is ready to be connected to the Keysight OMA Software as described in the following section. Keysight M8290A Getting Started Guide...
  • Page 32 Using the Instrument Setting up the M829xA in the Keysight OMA Software A standard M8290A system is pre-configured at the factory. This section shows how to modify the factory configuration or to configure a stand-alone module. In the Keysight 89600 VSA Software interface, select Utilities >...
  • Page 33 Using the Instrument From the Possible Logical Instruments list, select Keysight M8290 Optical Modulation Analyzer then click on the button to add it to the Configuration list. Keysight M8290A Getting Started Guide...
  • Page 34 Using the Instrument In the ADC selection, click on the down arrow and select the desired instrument. Keysight M8290A Getting Started Guide...
  • Page 35 In the Synthesizer selection, click on the down arrow and select the LO configuration (internal or external LO). With the M8296A only an external manually controlled LO is supported. This setting can also be used if the test setup does not actually include an LO. Keysight M8290A Getting Started Guide...
  • Page 36 Using the Instrument In the name field, enter the desired name for this configuration then click on the OK button. Repeat this procedure to add more configurations. Keysight M8290A Getting Started Guide...
  • Page 37: Esm Front Panel Connector

    Keysight M8290A Optical Modulation Analyzer and High-Speed Digitizer Test Solution Getting Started Guide AXle Chassis / 37 ESM Front Panel Connector / 39 Setting the Startup Delay for the Embedded Controller This chapter describes the usage of the AXIe chassis in combination with the M829xA.
  • Page 38 Connects a host PC to the chassis via USB 2.0. The USB 2.0 port is only available for AXIe chassis with option -U20. Using USB to connect to the M8290A is not supported. Multiframe Input Synchronizes timing signals with multiple daisy-chained chassis.
  • Page 39 M829xA module to work properly, the AXIe modules must boot in a certain order. To assure that this order is maintained at all times, a startup delay needs to be configured on the AXIe chassis’ system module. All M8290A systems are pre-configured accordingly at the factory. However, if a different AXIe chassis is used with any M829xA module, it may be necessary to set the startup delay manually.
  • Page 40 AXle Chassis Click on the “Modules” button ( ) on the left side: Keysight M8290A Getting Started Guide...
  • Page 41 Browse to the bottom of the “Chassis Module Configuration” box: If the “Active Configuration” shows a wait time below 15 seconds, select “Auto” in the “New Configuration” list and enter 15 seconds as the wait time. Then press “Apply New PCIe Settings”. Keysight M8290A Getting Started Guide...
  • Page 42 AXle Chassis Shutdown the embedded controller and power cycle the AXIe chassis. This setting is now shown as the “Active Configuration”: Keysight M8290A Getting Started Guide...
  • Page 43: Introduction

    Keysight M8290A Optical Modulation Analyzer and High-Speed Digitizer Test Solution Getting Started Guide M829xA Maintenance / 43 Introduction / 44 Initial Safety Information / 46 ESD Protection / 47 Power and Ventilation Requirements / 47 Thermal Protection / 47 Battery...
  • Page 44 Max. CW output power is defined as the highest possible optical CW power that the laser source can produce at its output. † Max. permissible CW output power is defined as the highest optical power that is permitted within the appropriate IEC laser class. Keysight M8290A Getting Started Guide...
  • Page 45: Laser Safety Labels

    IEC 60825-1 we recommend that you stick the laser safety labels, in your language, onto a suitable location on the outside of the instrument where they are clearly visible to anyone using the instrument. Keysight M8290A Getting Started Guide...
  • Page 46: Esd Protection

    Be sure the instrument and all connected devices (DUT, etc.) are properly earth-grounded (to a common ground) to prevent buildup of static charge and electrical over-stress. Keysight M8290A Getting Started Guide...
  • Page 47: Power And Ventilation Requirements

    When the temperature exceeds approximately 80°C the overheating detection switches off the instrument for safety reasons. For reliability, it is recommended to send instruments with broken or defective fans immediately to Keysight Service for repair. Battery The M829xA does not have a battery.
  • Page 49: M8292A Modular Optical Modulation Analyzer Specifications

    Keysight M8290A Optical Modulation Analyzer and High-Speed Digitizer Test Solution Getting Started Guide Specifications / 50 M8292A Modular Optical Modulation Analyzer Specifications / 53 M8296A Electrical High-Speed Digitizer Specifications / 55 General Characteristics / 55 Declaration of Conformity...
  • Page 50 Wavelength settling time < 30 s Wavelength uncertainty ±4.5 pm (±560 MHz), guaranteed ±2.5 pm (±310 MHz), typical Sidemode Suppression Ratio (SMSR) ≥50 dB Relative Intensity Noise (RIN) –145 dB/Hz (10 MHz to 40 GHz) Keysight M8290A Getting Started Guide...
  • Page 51 40 Gbaud symbol rate Table 5 M8292A Modular Optical Modulation Analyzer specifications (Continued) Trigger input Typical values Input range –4 V to +4 V Threshold –4 V to +4 V Range 10 mV Resolution 100 mV Keysight M8290A Getting Started Guide...
  • Page 52 Table 6 M8292A maximum ratings Maximum ratings Maximum signal input power +14.5 dBm Signal input damage level +15 dBm External LO input power Maximum +17 dBm Damage level +18 dBm Reference clock input damage level Keysight M8290A Getting Started Guide...
  • Page 53 Number of input channels Trigger input Input range –4 V to +4 V Threshold Range –4 V to +4 V Resolution 10 mV Sensitivity 100 mV Polarity Selectable: positive, negative, either edge Timing uncertainty ≤ 8 ns Keysight M8290A Getting Started Guide...
  • Page 54 Adjusted baseband frequency range available for signal analysis. Determined from a 9 order polynomial fit to the measured amplitude response. Table 8 M8296A maximum ratings Maximum ratings Damage level (single ended) 0.9 V / 3 VDC Damage level Keysight M8290A Getting Started Guide...
  • Page 55 EMC tested to IEC61326-1 Form factor 2-slot AXIe 1-slot AXIe Warm-up time 30 minutes Recommended 2 years re-calibration interval Declaration of Conformity Click the following link to view or download the latest version of DoC: http://www.keysight.com/go/conformity Keysight M8290A Getting Started Guide...
  • Page 56 This information is subject to change without notice. © Keysight Technologies 2017 Edition 1.0, November 2017 www.keysight.com...