3. If the frequencies are set properly, use the
Telemetry Service Tool or Wave Viewer to check
the transmitter information. Refer to NO SIGNAL
INOP and an RF OUT OF LOCK INOP at
Telemetry Service Tool or Wave Viewer .
4. Make certain that the patient is in the antenna
coverage area.
5. Check if the leadset is attached (the leadset is the
broadcast antenna). Leadset wires must be
shielded. The RF connection is made only to the
shields.
6. The transmitter is not transmitting a signal. This
can be a result of a dead battery or a power supply
failure within the transmitter. To determine if this
is the problem, remove one of the leads from the
patient and see if the corresponding Leads Off LED
illuminates. If it does, proceed to Step 7. If it does
not, do the following:
a. Make certain that the battery is installed properly.
b. Replace the battery with a fresh one.
c. If an ECG and SpO
used (even if SpO
certain that a lithium or alkaline battery is being
used). A zinc-air battery cannot be used with an
ECG and SpO
SpO
the battery compartment. The option string "1SP"
indicates that the SpO
d. Open the transmitter and make sure that the RF
Cable is connected to the Front End Assembly and
to the Main PCB. If not, connect it properly using
the RF Cable seating tool following the procedures
for installing the RF Cable in Chapter 6 of the
Service and Reference Guide.
e. Replace the Main PCB in the transmitter.
26
capable transmitter. To check if the
2
option is installed in the transmitter, look in
2
capable transmitter is being
2
is not being monitored) make
2
option is installed.
2
RF Troubleshooting Procedures