In this example, you can see that the Measure Gate includes only five rising edges. Therefore,
parameter calculations for rise time are performed only on the five pulses bounded by the gate
posts. The position of the gate posts is shown in the Start and Stop fields in the accompanying
dialog.

To Set Up Measure Gate

1. In the menu bar, touch Measure Setup...
2. Touch the Px tab for the parameter you want to gate. A mini-dialog to the right of the main
setup dialog opens.
Note: If you already have the parameter of interest set up, you can simply touch the parameter directly below the grid:
Example Parameter Readout
3. Touch inside the Start data entry field and enter a value, using the pop-up numeric keypad.
Or, you can simply touch the leftmost grid line and drag the gate post to the right.
4. Touch inside the Stop data entry field and enter a value, using the pop-up numeric keypad.
Or, you can simply touch the rightmost grid line and drag the gate post to the left.
WM-OM-E Rev I
X-Stream Operator's Manual
175