WARNING:
WARNING:
WARNING:
Caution:
Caution:
12
Chapter 3: Safety
• Do not connect the system power supply or docking system to a multiple
portable socket outlet (MPSO) or extension cord.
• Before using the transducer, inspect the transducer face, housing, and cable.
Do not use the transducer if the transducer or cable is damaged.
• Always disconnect the power supply from the system before cleaning the
system.
• Do not use any transducer that has been immersed beyond the specified
cleaning or disinfection level. See
• Use only accessories and peripherals recommended by SonoSite, including
the power supply. Connection of accessories and peripherals not
recommended by SonoSite could result in electrical shock. Contact SonoSite
or your local representative for a list of accessories and peripherals available
from or recommended by SonoSite.
To avoid the risk of electrical shock and fire hazard:
• Inspect the power supply, AC power cords, cables, and plugs on a regular
basis. Ensure that they are not damaged.
• The power cord set that connects the power supply of the ultrasound system
or the stand to mains power must only be used with the power supply or
docking system, and cannot be used to connect other devices to mains
power.
To prevent injury to the operator/bystander, the transducer must be removed
from patient contact before the application of a high-voltage defibrillation pulse.
To avoid possible electrical shock or electromagnetic interference, verify proper
operation and compliance with relevant safety standards for all equipment
before clinical use. Connecting additional equipment to the ultrasound system
constitutes configuring a medical system. SonoSite recommends verifying that
the system, all combinations of equipment, and accessories connected to the
ultrasound system comply with JACHO installation requirements and/or safety
standards such as AAMI-ES1, NFPA 99 OR IEC Standard 60601-1-1 and
electromagnetic compatibility standard IEC 60601-1-2 (Electromagnetic
compatibility), and are certified according to IEC Standard 60950 (Information
Technology Equipment (ITE)).
Do not use the system if an error message appears on the LCD display. Note
the error code and call SonoSite Technical Support for further assistance.
To avoid increasing the system and transducer connector temperature, do not
block the airflow to the ventilation holes on the side of the system.