Calibration Intervals (Int 3, Int 4) - wtw pH/Oxi 340i Operating Manual

Ph/oxygen measuring instrument
Hide thumbs Also See for pH/Oxi 340i:
Table of Contents

Advertisement

pH/Oxi 340i
Setting the calibration
interval
4.5

Calibration intervals (Int 3, Int 4)

For each measured variable, a time interval is stored. When it has ex-
pired, you will be reminded to calibrate. After a calibration interval has
expired, the probe symbol of the relevant measured variable flashes. It
is still possible to measure. By calibrating the relevant probe, the func-
tion is reset and the interval starts anew.
The following calibration intervals are set in the factory:
Measured
Designation
parameter
pH/ORP
Int 3
Dissolved oxy-
Int 4
gen
You can change each of these intervals (1 ... 999 days):
1
Switch off the measuring instrument.
2
Press the <M> key and hold it down.
3
Press the <ON/OFF> key.
The display test appears briefly on the display. After this, the
measuring instrument automatically switches over to configu-
ration.
4
Press the <RUN/ENTER> key repeatedly until Int 3 or Int 4 and
the required measured variable pH or O
TDS
pH
O
S
n
t 3
1
Time
Baud
d
1
8
Day.Month No.
Year
Ident
Tref25 Tref20
AutoCal DIN
LoBat
nLF
Lin
Cal
AutoCal TEC
5
Set the required interval (in days) until the next calibration with
< > < >.
6
Confirm with <RUN/ENTER>.
7
Switch to the measuring mode with <M>.
Default setting
7 days (d 7)
14 days (d 14)