Entering The Salt Content (Salinity) - wtw pH/Oxi 340i Operating Manual

Ph/oxygen measuring instrument
Hide thumbs Also See for pH/Oxi 340i:
Table of Contents

Advertisement

Operation
Entering the salinity
40
Note
In chapter 6 W
HAT TO DO IF
to take for error elimination.

4.4.6 Entering the salt content (salinity)

A salt content correction is required in the oxygen concentration mea-
surement of test samples with a salt content of more than 1 g/l. To do
this, you have to enter the salinity equivalent of the test sample (range
0.0 - 70.0) and to switch on the salinity correction.
1
Determine the salinity of the test sample.
2
Press the <CAL> key repeatedly until Sal appears on the dis-
play.
TDS UpH
O
S
1
5
5
0
Time
Baud
6
8
9
Day.Month No.
Year
Ident
Tref25 Tref20
AutoCal DIN
LoBat
nLF
Lin
Cal
AutoCal TEC
3
Enter the salt content with < > < >.
4
Switch to the measuring mode with <M>.
Note
How to switch on the salt content correction is described on page 34.