Operation
Criteria
36

4.4.4 AutoRead AR (Drift control)

The AutoRead function (drift control) checks the stability of the mea-
surement signal. The stability has a considerable impact on the repro-
ducibility of the measured values.
1
Call up the measuring mode with <M> and/or < > < > .
2
Activate the AutoRead function with <AR>. The current mea-
sured value is frozen (hold function).
3
Start AutoRead with <RUN/ENTER>.
AR flashes until a stable measured value is reached. This mea-
sured value is transmitted to the interface.
TDS
U
pH
O
S
1
7
9 2
Baud
Time
8
1
7
Day.Month No.
Year
Ident
Tref25
Tref20
AutoCal DIN
LoBat
nLF
Lin
Cal
AutoCal TEC
4
If necessary, start the next AutoRead measurement with
<RUN/ENTER>.
5
To terminate the AutoRead function: Press the <AR> key.
With identical measurement conditions, the following applies:
Measured
parameter
D. O. concentra-
tion
D. O. saturation
index
Sal
mV
/pH
%
mg/l
m
S/
cm
mbar
M
cm
1/ cm Sal
6
°
C
/
%
K
TP
°
F
Auto Store
ARng
AR
RCL
Reproducibility
better than 0.05 mg/l
better than 0.6 %
pH/Oxi 340i
Response time
> 10 seconds
> 10 seconds