What to do if...
Measured value too low
Measured value too
Display LoBat
Instrument does not
to keystroke
Display
Probe symbol flashes
Message StoFull
StoFull
StoFull
StoFull
66
Cause
– Insufficient flow
Cause
high
– High amount of dissolved
substances
– Air bubbles bump on the
membrane with high velocity
– The carbon dioxide pressure is
too high (> 1 bar)
6.3
General errors
Cause
– Batteries almost empty
Cause
react
– Operating condition undefined
or EMC load unallowed
to
to
to
to
Cause
– Time-out of the interface
Cause
– Calibration interval expired
Cause
– All memory locations are full
Remedy
– Provide flow to the probe
Remedy
– Correct solubility function
using the salinity equivalent
– Avoid direct flow to the
membrane
– Measuring not possible
Remedy
– Exchange the batteries (see
section 5.1 M
AINTENANCE
Remedy
– Processor reset:
Switch the instrument on
while pressing the <CAL> key
Remedy
– Check the instrument that is
connected
Remedy
– Recalibrate the measuring
system
Remedy
– Output data storage and clear
data storage
pH/Oxi 340i
)