Ph Calibration - wtw pH/Oxi 340i Operating Manual

Ph/oxygen measuring instrument
Hide thumbs Also See for pH/Oxi 340i:
Table of Contents

Advertisement

pH/Oxi 340i
Why calibrate?
When to calibrate?
AutoCal TEC
AutoCal DIN
ConCal
AutoRead
Displaying the calibra-
tion data
Printing the calibration
record
4.3

pH calibration

pH electrodes age. This changes the asymmetry (zero point) and slope
of the pH electrode. As a result, an inexact measured value is dis-
played. Calibration determines the current values of the asymmetry
and slope of the electrode and stores them in the measuring instru-
ment. Thus, you should calibrate at regular intervals.
l After connecting another electrode
l When the sensor symbol flashes
– (after the calibration interval has expired)
– after a voltage interruption, e. g. after changing the batteries
You can select one of 3 calibration procedures:
is specially matched to the WTW technical buffer solutions as a fully au-
tomatic two-point calibration. The buffer solutions are automatically
recognized by the measuring instrument. Depending on the instrument
setting (see section 4.8 C
relevant buffer nominal value or the current electrode voltage in mV.
The calibration can be terminated after the first buffer solution. This cor-
responds to a single-point calibration. When doing so, the instrument
uses the Nernst slope (-59,2 mV/pH at 25 °C) and determines the
asymmetry of the electrode.
is specially adapted to the permanently programmed buffer solutions in
accordance with DIN 19266 as a fully automatic two-point calibration.
The operating sequence of the AutoCal DIN calibration corresponds to
that of the AutoCal TEC calibration. The calibration can only be termi-
nated after the first buffer solution (single point calibration).
This function is a conventional two-point calibration using two buffer
solutions (pH 7.0 ± 0.5 and any other buffer solution) or a single-point
calibration using any buffer solution, which is used as a high-speed
method.
When calibrating with AutoCal TEC and AutoCal DIN, the AutoRead
function is automatically activated.
The current AutoRead measurement can be terminated at any time
(accepting the current value) by pressing <RUN/ENTER>.
You can view the data of the last calibration on the display. The pro-
ceeding is described on page 48.
The calibration protocol contains the calibration data of the current cal-
ibration. You can transmit the calibration protocol to a printer via the se-
rial interface (see page 50).
), the instrument displays the
ONFIGURATION
Operation
21

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents