wtw pH/Oxi 340i Operating Manual page 31

Ph/oxygen measuring instrument
Hide thumbs Also See for pH/Oxi 340i:
Table of Contents

Advertisement

pH/Oxi 340i
TDS
U
pH
O
S
1
S
L O
Time
Baud
2
8
4
Day.Month No.
Year
Ident
Tref25
Tref20
AutoCal DIN
LoBat
nLF
Lin
Cal
AutoCal TEC
8
Thoroughly rinse the electrode with deionized water.
9
If required, enter the temperature of the second buffer solution
with < > < > while keeping the <RUN/ENTER> key de-
pressed.
10
Immerse the pH electrode in the second buffer solution.
11
Press the <RUN/ENTER> key.
12
Set the measured value to the nominal pH value of the buffer
solution (at the current temperature) with < > < >.
13
When the measured value is stable, press the <RUN/ENTER>
key. The value of the slope (mV/pH) appears on the display.
The probe symbol shows the evaluation of the current calibra-
tion.
TDS
U
pH
O
S
1
5
9 2
Baud
Time
S
L
Day.Month
No.
Year
Ident
Tref25
Tref20
AutoCal DIN
LoBat
nLF
Lin
Cal
AutoCal TEC
14
Press the <RUN/ENTER> key. The value of the asymmetry
(mV) appears on the display.
Sal
mV/pH
% mg/l
m
S/
cm
mbar
M
cm
1/ cm
Sal
8
°
C
/
%
K
TP
°
F
Auto Store
ARng
AR
RCL
Sal
mV/pH
%
mg/l
m
S/
cm
mbar
M
cm
1/
cm
Sal
O
°
C
/
%
K
TP
°
F
Auto
Store
ARng
AR
RCL
Operation
31