Welch Allyn RetinaVue 100 Imager Directions For Use Manual page 47

Hide thumbs Also See for RetinaVue 100 Imager:
Table of Contents

Advertisement

Directions for use
Error message/problem
Camera is docked. Please lift the camera before saving, deleting
or starting an acquisition.
The battery is running low. Please recharge.
The battery level is too low to acquire any image. Please recharge
now.
Troubleshooting 43
Solution
Remove the camera from the docking station.
When the battery charge level is between eleven
and twenty-five percent, the low battery error
message appears. Place the device into the docking
station to charge the battery.
When the battery charge level is below ten percent
and the battery icon blinks, the low battery error
message appears. Place the device into the docking
station to charge the battery.