Holding Registers; Read Input Register; A.3 Holding Registers - Kipp & Zonen SMP series Instruction Manual

Smart pyranometer
Hide thumbs Also See for SMP series:
Table of Contents

Advertisement

.

A.3 Holding registers

Device Control
PDU address
Parameter
34
IO_DEF_SCALE_FACTOR
35 to 40
Factory use only
A.4 Read input register
Many of the registers and controls are for remote diagnostics. In this chapter only the most interesting registers and controls
are described.
Register 0
IO_DEVICE_TYPE
The device type defines which device is connected. This register can be used to check the type of the connected device.
IO_datamodel_version 102 supports the following type of sensors.
Real-time Processed Data
Parameter name
Register R/W
IO_DEVICE_TYPE
R0
Parameter
Value
SMP3
601
(volt version)
SMP3
602
(current loop version)
SMP6
619
(volt version)
SMP6
620
(current version)
SMP10
617
(volt version)
SMP10
618
(current version)
SMP11
603
(volt version)
SMP11
604
(current loop version)
SMP21
605
(volt version)
SMP21
606
(current loop version)
SMP22
607
(volt version)
SMP22
608
(current loop version)
SGR3
609
(volt version)
SGR3
610
(current loop version)
SGR4
611
(volt version)
SGR4
612
(current loop version)
SHP1
613
(volt version)
SHP1
614
(current loop version)
SUV5
615
(volt version)
SUV5
616
(current loop version)
R/W
Type
Mode
R/W
S16
All
Initial Val
R
65535
# of sensors
1/Sensitivity
1
5-20 V/(W/m²)
1
5-20 V/(W/m²)
1
7-14 V/(W/m²)
1
7-14 V/(W/m²)
1
7-14 V/(W/m²)
1
7-14 V/(W/m²)
1
7-14 V/(W/m²)
1
7-14 V/(W/m²)
1
7-14 V/(W/m²)
1
7-14 V/(W/m²)
1
7-14 V/(W/m²)
1
7-14 V/(W/m²)
2*
5-15 V/(W/m²)
2*
5-15 V/(W/m²)
2*
5-15 V/(W/m²)
2*
5-15 V/(W/m²)
1
7-14 V/(W/m²)
1
7-14 V/(W/m²)
1
300 - 500 V/(W/m²)
1
300 - 500 V/(W/m²)
Description
Default scale factor
Mode
Description
All
Selected device type of the sensor
Type
Pyranometer
Pyranometer
Pyranometer
Pyranometer
Pyranometer
Pyranometer
Pyranometer
Pyranometer
Pyranometer
Pyranometer
Pyranometer
Pyranometer
Pyrgeometer
Pyrgeometer
Pyrgeometer
Pyrgeometer
Pyrheliometer
Pyrheliometer
UV Radiometer
UV Radiometer
43

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Smp6Smp10Smp3Smp11Smp21Smp22