8. Trouble shooting

.
There are no user-serviceable parts within the SMP pyranometer and it must not be opened without the agreement and instruction
of Kipp & Zonen.

8.1 Output signal not present or incorrect

The following contains a procedure for checking the instrument in case it appears that it does not function correctly:
1. Check the SMP pyranometer cable wires are properly connected to the readout equipment.
2. Check the power supply (12 VDC recommended).
3. Check that the instrument has a unique Modbus® address.
4. Compare the digital and analogue outputs to see if the problem is in one output only.
5. Check the instrument location. Are there any obstructions that cast a shadow on the instrument by blocking the direct sun
during some part of the day?
6. Check the dome, it should be clear and clean. If condensation is deposited on the inside, please change the desiccant. If too
much water is deposited internally the drying cartridge should be removed and the instrument warmed to dry it and then
replace with new desiccant. It may take several days for the sensitivity to fully recover to the original value.
7. For analogue outputs check the data logger or integrator input o set such that a signal of 0 Volt or 4 mA (as appropriate)
gives a 'zero' reading.
8. Check levelling. The bubble should be at least half inside the marked ring of the level.
9. If water, frost or ice is deposited on the dome, clean it. Probably water droplets will evaporate in less than one hour under
sunlight.
Any malfunction or visible damage should be reported to your Kipp & Zonen representative, who will suggest the appropriate action.

8.2 Frequently Asked Questions

The most frequently asked questions are listed below. For an update or further information refer to our website at
www.kippzonen.com
Q: Negative output during night-time measurements?
A: This error is related to Zero O set Type A. Normally this zero o set is present when the inner dome has a di erent temperature
from the cold junctions of the sensor (the instrument housing). Practically this is always the case when there is a clear sky. Because
of the low e ective sky temperature (< 0°C) the Earth's surface emits roughly 100 W/m² of long-wave infrared radiation upwards.
The outer glass dome of a pyranometer also has this emission and is cooling down several degrees below air temperature (the
emissivity of glass for the particular wavelength region is nearly 1).
The emitted heat is attracted from the body by conduction in the domes, by wind and from the domes through infrared radiation.
The heat flow is opposite to the heat flow when absorbing solar radiation and causes the well-known zero depression at night.
This negative zero o set is also present on a clear day but is hidden in the solar radiation signal.
Zero O set Type A can be checked by placing a light and IR reflecting cap over the pyranometer. The response to solar radiation
will decay with the response time of the instrument, but the dome temperature will go to equilibrium with a time constant of
several minutes. So after about half a minute the remaining signal is mainly Zero O set Type A.
Good ventilation of the domes and housing minimises zero o sets and increases stability. Using the Kipp & Zonen CVF4 ventilation
unit can reduce Zero O set Type A by about 50 %.