System Circuit Breaker; Elapsed Time Meter; Control Circuit Breaker; Power On Indicator - HELIX 9600 Installation, Operation And Maintenance Instructions

High voltage compressor
Table of Contents

Advertisement

System Circuit Breaker

Elapsed Time Meter

Control Circuit Breaker

Power ON Indicator

Gas Charge Flared Fitting

©2007 Helix Technology Corporation
The System Circuit Breaker protects main input power to the Compressor
pump and module. The circuit breaker positions are labeled ON (1), which
is in the UP position, and OFF (0), which is in the DOWN position.
NOTE: The phase monitor in the Compressor will cause the system circuit
breaker to open when input power phases are incorrect.
The Elapsed Time Meter records the number of Compressor operating
hours. Since the meter is digital, it is not illuminated unless the system
circuit breaker is in the ON position and power is connected to the
Compressor. The Elapsed Time Meter maintains the correct accumulated
operating hours while system power is turned OFF.
NOTE: The meter cannot be reset.
The Control Circuit Breaker provides current overload protection for all
internal components of the Compressor except the Compressor motor. The
Compressor motor is protected by a separate overload protector. The
Control Circuit Breaker opens automatically and must be reset manually.
The Power On Indicator illuminates when the system circuit breaker is
placed in the ON position. The Compressor pump is energized when the
power indicator is illuminated and the elapsed time meter records system
operation time.
The Gas Charge Flared Fitting is used to connect a 99.999% pure helium
supply to the Compressor when helium charging is required. The fitting has
a 45 º flare and 7/16 in. x 20 threads/inch. Refer to
Maintenance"
for information on adding helium to the Compressor.
Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068
"Section 5 -
1-9