10. Disinfect the video baton according to the conditions stated in
vary depending on the solution and the component.
IMPORTANT
If you are using pre‑moistened wipes, wipe down the component as often as needed to keep it
visibly wet throughout the specified exposure time. You may use multiple wipes if necessary.
11. If applicable, rinse the video baton according to the solution manufacturer's instructions.
12. Dry the video baton as appropriate for the disinfectant you are using.
If you are using ...
A liquid disinfectant
Pre‑moistened wipes
13. Inspect the video baton according to the instructions in the following procedure, and then store the
disinfected video baton in a clean environment.
PROCEdURE 7.
WARNING
Before every use, ensure the instrument is operating correctly and has no sign of damage. Do
not use this product if the device appears damaged. Always ensure that alternative airway
management methods and equipment are readily available.
Report any suspected defects to Verathon
verathon.com/support.
1. Visually inspect the video baton for signs of damage. Perform a routine inspection of the video baton
before and after every use to ensure that all endoscopic components are free of unintended rough
surfaces, sharp edges, protrusions, or cracks.
Operations & Maintenance Manual: Cleaning & Disinfecting
Then ...
Use a sterile cloth, hospital‑grade clean air, or a low‑temperature dryer
to dry the component.
Allow the component to air dry completely.
INSPECT THE VIdEO BATON
Table
8. The exposure process and times
Customer Care. For contact information, visit
®
45