Operations & Maintenance Manual: Cleaning & Disinfecting
DISINFECTION
CYCLES*
LEVEL
High
750
High
1500
High
1500
High
1500
High
1500
High
1500
High
1500
CONDITIONS
Standard cycles in the following processors:
STERIS
SYSTEM 1
(outside U.S.)
®
®
SYSTEM 1E
(in U.S.)
®
SYSTEM 1 EXPRESS (outside U.S.)
SYSTEM 1 PLUS (outside U.S.)
Clean according to the entry
System"
in
Table 9
on page 46.
Sporocidal: Apply two pumps of the activator foam
to a sporocidal towelette and manipulate towelette for
15 seconds. Wet all surfaces of the component and allow
to remain wet for 30 seconds.
Rinse: Use a rinse towelette to wipe all surfaces.
Conditioning: 20°C (68°F) or higher
Water Temperature: 20°C (68°F) or higher
Exposure: Soak for 8 minutes, ensuring that all air
bubbles are removed from the surface of the component.
Rinse: (1) 1‑minute immersion with agitation in
pure water. Ensure the connector is properly rinsed.
Conditioning: 20°C (68°F) or higher
Water Temperature: 20°C (68°F) or higher
Exposure: Soak for 10 minutes, ensuring that all air
bubbles are removed from the surface of the cable.
Rinse: (3) 1‑minute immersions with agitation in pure water.
Conditioning: 20°C (68°F) or higher
Water Temperature: 20°C (68°F) or higher
Exposure: Soak for 10 minutes, ensuring that all air
bubbles are removed from the surface of the cable.
Rinse: (3) 1‑minute immersions with agitation in pure water.
Conditioning: 20°C (68°F) or higher
Water Temperature: 20°C (68°F) or higher
Exposure: Soak for 10 minutes, ensuring that all air
bubbles are removed from the surface of the blade.
Rinse: (3) 1‑minute immersions with agitation in pure water
Water Temperature: 20°C (68°F)
Exposure: Soak the component for 30 minutes in
OPASTER solution, ensuring that all air bubbles are
removed from the surface. Use the solution at full
strength.
Rinse: (3) 1‑minute immersions with agitation in pure
water. Ensure that any exposed connectors are properly
rinsed.
"Tristel™ Trio Wipes
51