Cleaning Information

How to clean connector interfaces

CAUTION
92
Preferred Procedure
Use the following procedure on most occasions.
1
Clean the connector by rubbing a new, dry cotton-swab over the
surface using a small circular movement.
2
Blow away any remaining lint with compressed air.
Procedure for Stubborn Dirt
Use this procedure particularly when there is greasy dirt on the connector:
1
Moisten a new cotton-swab with isopropyl alcohol.
2
Clean the connector by rubbing the cotton-swab over the surface
using a small circular movement.
3
Take a new, dry soft-tissue and remove the alcohol, dissolved sediment
and dust, by rubbing gently over the surface using a small circular
movement.
4
Blow away any remaining lint with compressed air.
An Alternative Procedure
A better, more gentle, but more expensive cleaning procedure is to use an
ultrasonic bath with isopropyl alcohol.
1
Hold the tip of the connector in the bath for at least three minutes.
2
Take a new, dry soft-tissue and remove the alcohol, dissolved sediment
and dust, by rubbing gently over the surface using a small circular
movement.
3
Blow away any remaining lint with compressed air.
Be careful when using pipe-cleaners, as the core and the bristles of the
pipe-cleaner are hard and can damage the interface.
Do not use pipe-cleaners on optical head adapters, as the hard core of
normal pipe cleaners can damage the bottom of an adapter.