Endress+Hauser
Divisible probes
If there is little mounting space (distance to the ceiling), it is advisable to use divisible rod
probes ( 16 mm).
• max. probe length 10 m (394 in)
• max. sideways capacity 30 Nm
• probes are separable several times with the following lengths of the individual parts:
– 500 mm (20 in)
– 1 000 mm (40 in)
• torque: 15 Nm
14 mm
14 mm
Mounting
A0021647
25