Operating menu
Factory setting
Additional information
Value current output 1 to 2
Navigation
Prerequisite
Description
User entry
Factory setting
Additional information
Switch output simulation
Navigation
Description
Selection
Factory setting
Switch status
Navigation
Prerequisite
Description
Selection
Factory setting
230
Off
An active simulation is indicated by a diagnostic message of the Function check (C)
category.
Expert → Diagnostics → Simulation → Value current output 1 to 2
Current output simulation (→ 229) = On
Defines the value of the simulated output current.
3.59 to 22.5 mA
3.59 mA
The current output assumes the value specified in this parameter. In this way, users can
verify the correct adjustment of the current output and the correct function of connected
control units.
Expert → Diagnostics → Simulation → Switch output simulation
Switch the simulation of the switch output on and off.
• Off
• On
Off
Expert → Diagnostics → Simulation → Switch status
Switch output simulation (→ 230) = On
Current status of the switch output.
• Open
• Closed
Open
Levelflex FMP51, FMP52, FMP54 HART
Endress+Hauser
Need help?