Levelflex FMP51, FMP52, FMP54 HART
Assign measurement variable
Navigation
Selection
Factory setting
Additional information
Process variable value
Navigation
Prerequisite
User entry
Factory setting
Additional information
Current output 1 to 2 simulation
Navigation
Description
Selection
*
Visibility depends on order options or device settings
Endress+Hauser
Description of parameters
Navigation
Expert → Diagnostics → Simulation
Expert → Diagnostics → Simulation → Assign measurement variable
• Off
• Level
*
• Interface
*
• Thickness upper layer
• Level linearized
• Interface linearized
• Thickness linearized
Off
• The value of the variable to be simulated is defined in the Process variable value
parameter (→ 229).
• If Assign measurement variable ≠ Off, a simulation is active. This is indicated by a
diagnotic message of the Function check (C) category.
Expert → Diagnostics → Simulation → Process variable value
Assign measurement variable (→ 229) ≠ Off
Signed floating-point number
0
Downstream measured value processing and the signal output use this simulation value. In
this way, users can verify whether the measuring device has been configured correctly.
Expert → Diagnostics → Simulation → Current output 1 to 2 simulation
Switch the simulation of the current output on and off.
• Off
• On
Operating menu
229