Command 152: Read Run-Time Meter; Command 154: Set Event Index - MSA FL500 UV/IR Operating Manual

Hart communication
Hide thumbs Also See for FL500 UV/IR:
Table of Contents

Advertisement

9.15. Command 152: Read Run-time Meter

Request Data Bytes
Byte
Format
0
N/A
Response Data Bytes
Byte
Format
0-3
Unsigned-32
Command-specific Response Codes
Code
Class
0
Success
17-127

9.16. Command 154: Set Event Index

Request Data Bytes
Byte
Format
0
Unsigned-8
Response Data Bytes
Byte
Format
0
Unsigned-8
Command-specific Response Codes
Code
Class
0
Success
1-2
3
Error
4
5
Error
6-127
Description
N/A
Description
Run-time meter value
Description
No command-specific errors
Undefined
Description
Sets the index event before reading the
event log
Range: 0 – 9
Description
Event index
Description
No command-specific errors
Undefined
Passed parameter too large
Undefined
Too few data bytes received
Undefined
FL500 HART® Communication Operating Manual
Device-specific Commands