1.
If replacing an Oxygen cartridge with another Oxygen cartridge ensure that the Oxygen
Transducer Adaptor is left in place on the Certified Sensor body.
2.
If replacing an Oxygen cartridge with a different type of cartridge ensure that the Oxygen
Transducer Adaptor is removed from the sensor body.
3.
If replacing a standard cartridge with an Oxygen cartridge ensure that the Oxygen Transducer
Adaptor supplied with the Oxygen cartridge is fitted to the Certified Sensor body. The adaptor
is fitted to the sensor body via a bayonet fitting.
The Change Cartridge screen now displays the following message.
(9)
Fit the replacement cartridge.
Caution: Only cartridges with the following part numbers can be fitted to the Certified
Sensor:
2110B30x0, 31x0, 32x0, 33x0, 34x0, 35x0 series
2110B3700 - 2110B3999 range
Notes:
1.
Certified to CSA 22.2 No. 152 only when fitted with specific cartridges.
See Specifications for details.
Carefully plug the cartridge into the Certified Sensor body ensuring that the cartridge tab
lines up with the recess in the sensor body and push the cartridge, without twisting, until it is
fully home.
If the cartridge will not push fully home, re-check that the locating tab is correctly aligned with
the recess in the sensor body. Position the tab so it is resting on the tab recess wall to the
right or left of the tab recess and then turn the cartridge until the tab drops into the recess.
Caution:
Do not force the cartridge as this may cause damage to the pins of the connecting
plugs. Twisting and pushing can bend the pins and make the cartridge inoperative.
5. MAINTENANCE
Change Cartridge
Change Cartridge Now
Cartridge: Not Present