Accessories - Honeywell Apex Technical Handbook

Gas detection system
Hide thumbs Also See for Apex:
Table of Contents

Advertisement

MAN0604_Issue 09_02-2013
2110M8030
APPENDIX A - SPECIFICATIONS
A.2.2 Catalytic SG16 cartridges, Methane and Propane
Range:
Operating Temperature Range:
Temperature Effects:
5%rH to 90%rH (99%rH
Humidity Effects:
Operating Pressure Range:
Pressure Effects:
Warm-up Time:
Response Time:
Linearity:
Accuracy/Repeatability:
Stability with Time (3 months):
Permitted Accessories:
2110B3701 - Propane, 0 to 100%LEL.
100%LEL eq. 1.7%V/V)
2110B3704 - Methane, 0 to 100%LEL.
(100%LEL eq. 4.4%V/V)
-40 to +65°C
2110B3704 – Methane, Better than +/-5% LEL
2110B3701 – Propane, +/-6%LEL
Operating Humidity Range:
Intermittent – non-condensing.)
Better than +/-5% LEL.
80kPa to 110kPa
Better than +/-5% LEL.
30s (typical). 900s (worst case).
2110B3701 – Propane, T
(typical)
2110B3704 – Methane, T
(typical)
(When fitted with standard sensor cap
and no filter.)
Better than +/-5% LEL or +/-10%
reading, whichever is the greater.
Better than +/-3% LEL
2110B3704 – Methane,
Better than +/-5% LEL
2110B3701 – Propane, +/-6%LEL
Standard sensor cap, plus either,
no filter of Mesh filter.
Weather housing, plus either, no filter,
Mesh filter, or Hydrophobic filter.
(Response Time will be increased when
using the Weather Housing and either
the Mesh or Hydrophobic Filters)
Collecting Cone, plus either, no filteor
Mesh filter.
Sun Shade
108
Apex
8s, T
17s.
50
90
6s, T
11s.
50
90

Hide quick links: