Table A.10: ADAM-4018/4018+ Range Accuracy for Thermocouple
Range Code (Hex)
0E
0F
10
11
12
13
14
Note!
Because the CJC sensor of ADAM-4018 is located in the side of chan-
nel 0 to 4, the measurement will have the difference ± 1 °C between
channel 0 ~ 4 and channel 5 ~ 7.
ADAM-4000 Series User Manual
Input Range
Typical Accuracy Maximum Error
J thermocouple
±1.0
0 to760 °C
K thermocouple
±1.0
0 to 1370 °C
T thermocouple
±1.0
-100 to 400 °C
E thermocouple
±1.0
0 to 1000 °C
R thermocouple
±1.2
500 to 1750 °C
S thermocouple
±1.2
500 to 1750 °C
B thermocouple
±2.0
500 to 1800 °C
294
Units
±1.5
°C
±1.5
°C
±1.5
°C
±1.5
°C
±2.5
°C
±2.5
°C
±3.0
°C