Sphh:mm" - Stop Logging; Sr" - Synchronize Readings - Geokon LC-2x4 Instruction Manual

4 channel vw datalogger
Table of Contents

Advertisement

32

3.32 "SPhh:mm" - Stop Logging

Stop the datalogger logging values; hh is the hour (24-hour format) of the day to stop and mm
the minutes. The time entry is optional.
*SC60
Scan interval: 60 second(s).
*ST
Logging started.
Datalogger#1,2007,11,25,14,10,05,2.94,23.99,9.071,---,---,
---,22.9,---,---,---,1
*SP12:00
Logging will start at: 15:13:46
Logging will stop at: 12:00:00
*
Note that when SPhh:mm is issued, the datalogger responds with the time of the next reading
along with the time at which logging will stop.

3.33 "SR" - Synchronize Readings

Synchronize readings to the top of the hour. If enabled (default) via the SR1 command, then all
readings after the first reading will synchronize to the top of the hour:
*SR1
Readings are synchronized to the top of the hour.
*ST
Logging started.
2008,318,1314,41,3.50,24.45,-8961.077,-999999.0,-999999.0,
-8444.892,23.1,-99.0,23.8,23.9,6645
*2008,318,1314,45,3.50,24.57,-8961.276,-999999.0,-999999.0,
-8445.080,23.2,-99.0,23.8,23.9,6646
2008,318,1315,0,3.50,24.86,-8960.023,-999999.0,-999999.0,
-8445.035,23.2,-99.0,23.8,23.9,6647
*SR0
Readings are not synchronized to the top of the hour.
*ST
Logging started.
2008,318,1316,31,3.50,24.39,-8960.209,-999999.0,-999999.0,
-8445.080,23.3,-99.0,23.8,23.9,6648
*2008,318,1316,46,3.50,24.80,-8960.090,-999999.0,-999999.0,
-8445.092,23.3,-99.0,23.8,23.9,6649
*2008,318,1317,1,3.50,24.80,-8961.173,-999999.0,-999999.0,
-8445.302,23.4,-99.0,23.8,23.9,6650