Log Intervals List
------------------------------------------------
Interval #1 Length: 10
Interval #2 Length: 20
Interval #3 Length: 30
Interval #4 Length: 40
Interval #5 Length: 50
Interval #6 Length: 60
*
The above list indicates that there are 96 iterations of interval #1 left before interval #2 begins
execution. See Section 3.12 to modify intervals.

3.12 "Ln/lllll/iii" - Define Length and Iteration of Interval

Define the length and iteration of any interval in the list; n refers to the number of the interval
(#1-#6), lllll is the length (3-86400), and iii is the iterations (0-255), or the number of readings
that will be taken at that interval. If 0 is entered for the iteration value that interval will execute
indefinitely. Illegal entries will be ignored (e.g., L7/10/100 or L1/1000/500). If the entry
is correct the modified interval will display.
*L1/100/0
Interval #1
*
If log intervals are enabled and logging was started, any change to the interval list will
result in a restart of the table!
Table 5 lists possible logarithmic interval lengths and iterations. Any combination of lengths and
iterations is permissible.
Interva
Iterations: 100/96
Iterations: 90/90
Iterations: 80/80
Iterations: 70/70
Iterations: 60/60
Iterations: 50/50
Length: 100
Length
l
1
10 seconds
2
30 seconds
3
60 seconds
4
300 seconds
5
2400 seconds
6
3600 seconds
Table 5 - Logarithmic Intervals List
Iterations: 0
Iterations
Elapsed Time
6
1 minute
20
10 minutes
100
100 minutes
200
1000 minutes
250
10000 minutes
0
endless