3. COMMAND LIST

3.1 "BRnnn"- Set or Display the Current Baud Rate

This command is not displayed in the help text (?<ENTER>). Displays or sets the current baud
rate. Valid numbers for "nnn" are 9 (9600 baud) or 115 (115,200 baud). The example below
changes the baud rate to 9600 with no confirmation.
*BR9

3.2 "C" - Display Current Clock Settings

Displays the current datalogger real-time clock settings. The CS command section explains how
to adjust the clock settings.
*C
Date: 02/21/07
*

3.3 "CSmm/dd/yy/hh:mm:ss" - Set the Internal Clock

Set the datalogger's internal real time clock; mm represents the month, dd the day of the month,
yy the year, hh the hours, mm the minutes, and ss the seconds. Illegal combinations will be
ignored (e.g., CS02/30/07 or CS///12:60). Fields that are left blank will not be changed (e.g.,
CS//07 to only change the year).
*CS///10:45:00
Date: 02/21/07
*
Note: If logging is currently started and the clock is changed, a restart of the scan interval or log
interval table will occur.
3.4 "DEFAULT" – Load Factory Default Settings
The DEFAULT command will reload the datalogger's channel and gage settings to the factory
default settings, along with the reading synchronization and memory wrap settings. This results
in:
All channels Enabled
All Gage Types set to 1
All Zero Readings set to 0.00000
All Gage Factors set to 1.00000
All Gage Offsets set to 0.00000
All channels use linear conversion
Scan interval = 10S
All readings synchronized to the top of the hour
Time: 10:43:08
Time: 10:45:00