The calibration protocol contains the calibration data of the
current calibration. You can call up the calibration protocol
by outputting the data storage (section 4.7.3 O
THE DATA STORAGE
Note
You can automatically print out a calibration protocol after
the calibration. To do so, switch on the printer before the cal-
ibration (Print LED lights up green). After a valid calibration,
the protocol is printed.
Sample printout:
CALIBRATION Protocol
02.03.99 / 16:20
Device No.: 1234
CALIBRATION ISE
Cal Time: 01.03.99 / 15:10
Std 1
10,01
mg/l
Std 2
20,00
mg/l
Std 3
50,00
mg/l
C1
0,5 mV
C2
8,4 mV
C3
21,2 mV
S1
26.4 mV
S2
32,1 mV
Probe:
+++
After the calibration, the instrument evaluates the current
status of the electrode, i. e. the slope of the electrode (mV).
The evaluation appears on the display.
Display
E3
).
25.0°C
24.9°C
25.0°C
Slope [mV]
50.0 ... 70.0
25.0 ... 35.0
Clear the fault according to
chapter 6 W
UTPUTTING
...
HAT TO DO IF