wtw inoLab pH/ION Level 2 Operating Manual page 29

Ph/ion meter with integrated printer
Hide thumbs Also See for inoLab pH/ION Level 2:
Table of Contents

Advertisement

Single-point
calibration
Only the electrode asymmetry is determined in single-point
calibration. The slope of the last two-point calibration is re-
tained.
Use a buffer solution in the range pH = 7.0 ± 0.5 for this pro-
cedure.
Note
Step 3 is not required if you use a temperature probe. The
TP message indicates an active temperature measurement.
m
1 Press the
the status line.
c
2 Press the
tion display appears.
3 Set the temperature of the buffer solution using
u d
while keeping the
4 Submerse the pH electrode in the buffer solution.
g
5 Press the
The measured pH value appears on the display.
6 Set the nominal pH value of the buffer solution (at the
current temperature) by pressing the
g
7 Press the
The value of the asymmetry (mV) and the sensor
symbol for the evaluation of the electrode appears on
the display.
8 To return to the measuring mode: Press the
key repeatedly until pH or U appears in
key repeatedly until the ConCal func-
g
key depressed.
key.
key.
Operation
u d
keys.
m
key.