Programming
Access P365 and enter EDIT mode
P365 Multispan
Select [1-8]
Span 1 will have already been set as part of the Start Up and initial calibration. Therefore,
select 2.
Access P017 and enter EDIT mode
P017 Test Rate: Weight MS2
Enter Test Rate
Enter the test rate value, and press
To do a span calibration for another condition, (i.e. span 3 or 4 etc.), access P365 and
repeat these steps for each condition. As with any initial span, follow the span calibration
for each multispan with a material test and factoring.
To use remote span selection, auxiliary inputs, 1 and/or 2 or 3, are programmed to read
the contact state as the span selection. Remote selection overrides the keypad (or
Dolphin Plus) selection. The auxiliary inputs override the keypad selection.
Access P270 and enter EDIT mode
P270-01 auxiliary Input Function
Select Function [0-13]
Enter
. This programs auxiliary Input 1 (terminal 24) to read the contact state for span
selections: 1 or 2.
Access P270 and enter EDIT mode (when using spans 3 and/or 4)
P270-02 auxiliary Input Function
Select Function [0-13]
Enter
. This programs auxiliary Input 2 (terminal 25), in conjunction with auxiliary
input 1 to read the contact state for span selections 3 and 4.
Access P270 and enter EDIT mode (when using spans 5 to 8)
P270-03 auxiliary Input Function
Select Function [0-13]
Enter
. This programs auxiliary input 3 (terminal 26), in conjunction with auxiliary
input 1 and auxiliary input 2 to read the contact state for span selections 5 to 8.
7ML19985CN01
Milltronics SF 500 – INSTRUCTION MANUAL
E
0
E
0
to do a span calibration.
E
0
E
0
E
0
Page 39