Start Up; Power Up; Programming - Siemens Milltronics SF 500 Instruction Manual

Table of Contents

Advertisement

Start Up

Notes:
For successful start up, ensure that all related system components such as the
flowmeter are properly installed and connected.
Ensure that the SW1 is in the normal position, to the right. (See Layout diagram on
page 7)
Initial start up of the SF 500 consists of several stages, and assumes that the physical and
electrical installation of the solids flowmeter is complete:

power up

programming

load cell balancing
zero and span calibration
Power Up
Upon initial power up, the SF 500 displays:
P001 Language
1-Eng
Programming
Press
The SF 500 then scrolls sequentially through the start up program as parameters P001
through P017 are addressed.
P002 Test Reference Selection
Select 1-Weight, 2-Ecal
Press
P003 Number of Load Cells
Enter (1 or 2)
Press
P004 Rate Measurement System
Select 1-Imperial, 2-Metric
7ML19985CN01
Milltronics SF 500 - INSTRUCTION MANUAL
V
The initial display prompts the user
to select the preferred language.
1
e.g. Accept 'weight' (supplied with
V
scale) as the test reference.
1
e.g. Accept '1' as the number of load
V
cells.
1
e.g. Accept '2' for measurements in
V
metric.
2
Page 23