7.2.1 Disassembly of 1000ul single channel pipettes
1. Remove the tip fitting 5 (Picture 8).
To remove the tip fitting place the star-shaped end of tool 3 (picture 9) into the end of the tip
fitting. Turn the tool counterclockwise to unscrew the tip fitting (picture 10).
5
Picture 8
2. Press the tip ejector 4.1. Rotate the tip ejector 10
counterclockwise and pull it out (Picture 11).
3. Turn the tip cone 11 counterclockwise with the service tool 1.
There are two slots in the service tool: the smaller one for
pipettes under 300 μl and the bigger one for pipettes over 300 μl.
4. Pull out the piston assembly and other parts from the handle
with the tip cone 11.
Then turn the tip cone upside down and tap all parts from it.
7.2.2 Disassembly of single channel pipettes 300ul and smaller

1. Press the tip ejector 4.1. Rotate the tip ejector 10 counterclockwise and pull it out (Picture 12).

NOTE! If you want to service only parts inside cylinder go to step 4. But then use the bigger slot in
service tool.
2. Remove the tip fitting 5 (Picture 8).
To remove the tip fitting place the star-shaped end of tool 3 (picture 9) into the end of the tip
fitting (tip fitting is removable by hand in models 0.1-2µl and 1-10 µl). Pull the Clip trigger 8
backwards slightly with the other hand.
Turn the tool counterclockwise to unscrew the
tip fitting (picture 10).
4.1
10
Picture 11
5.2
5.1
1
11
Picture 9
4.1
10
Picture 12
12
3
Picture 10
1
9
8
11