Change Control Cartridge; Change Sensor Card - NOVA BIOMEDICAL Stat Profile PRIME Instructions For Use Manual

Ccs analyzer
Table of Contents

Advertisement

Prime CCS Instructions for Use Manual

4.2.2 Change Control Cartridge

1. Press the Tool Box
2. From the System Menu, press the Change Control
Cartridge and wait for pump to stop.
3. Open the door, remove old cartridge.
4
Slide new cartridge in past the front retaining lip.
5. Close door, press the Calibrate

4.3 Change Sensor Card

1. Press the Tool Box
2. From the System Menu, press the Change Sensor
Card and wait for pump to stop.
3. Open the door, remove old cartridge.
4
Open Cartridge door. Insert new cartridge. Close
Cartridge door.
5. Close door, press the Calibrate
NOTE: Touch Sensor Card by the edges and replace it
the same way back as pictured.
Sensor Card
Hold from
Edges
4-4
Figure 4.4 Sensor Card
icon.
icon.
icon.
icon.