7. Use the sensor arm on the left of the sensor to push
the sensor to the right and off.
8. Use a syringe filled with distilled water and a tube
to attach to the S-line port.
9. Push the distilled water through the bottom of the
Reference Sensor.
10. Next attach the tubing of the syringe to the top right
pump tubing port. Push the distilled water through
the top of the Reference Sensor.
11. With the Reference Sensor cleared, take the sensor
by the arm and slide it back onto the analyzer from
right to left.
12. Plug the pump tubing manifold to the top of the
sensor.
13. Reconnect the S-line to the bottom of the sensor.
14. Replace the Sensor Card.
15. Close door, press the Calibrate

6.3.2 Flushing the Sensor Cartridge

1. Press the Tool Box
2. From the System Menu, press the Change Sensor
Card and wait for pump to stop.
3. Open the door, remove the Sensor Cartridge.
4. Use the Sensor Cartridge Flushing tool with syringe
to flush the cartridge.
5. Attach the Flushing Tool to the topnport of the
Sensor Cartridge and push distilled water through
the cartridge.
6
With the Sensor Cartridge now cleared, open
cartridge door, insert the cartridge, and close the
cartridge door.
5. Close the Prime door and press the Calibrate
icon.
NOTE: Touch Sensor Card by the edges and replace it
the same way back.
Appendix A
icon.
icon.
6-3