Prime CCS Instructions for Use Manual
5. Unplug pump tubing manifold from top of sensor.
6. Disconnect the S-line from the bottom of the sensor
using the removal tool attached to the door.
7. Use the Sensor Arm on the left of the sensor to
push the sensor to the right and off.
Slide
Reference
Sensor Off
Figure 5.9 Slide the Reference Sensor Off to the Right
8. Take a new sensor by the arm and slide it back onto
the analyzer from right to left.
9. Plug the pump tubing manifold to the top of the
sensor.
10. Reconnect the S-line to the bottom of the sensor.
11. Replace the Sensor Card. (See Figure 4.4.)
12. Close door, press the Calibrate
5-6
icon.