StatSensor Creatinine Control Solution
Read the StatSensor Creatinine Control Solution
package insert sheet for complete instructions, indi-
cations, precautions, and limitations of the system.
Only the Nova StatSensor Creatinine Control Solutions are
recommended for use with the Nova StatSensor Xpress-i
Creatinine Meter and the Nova StatSensor Creatinine Test
Strips. Ranges for the Nova StatSensor Xpress-i Creatinine
Meter using other commercially available creatinine controls
have not been established and may give erroneous results.
Run 2 different levels of the StatSensor Creatinine Control
Solutions during each 24 hours of testing prior to testing of
patient specimens and under the following circumstances:
Testing a Quality Control Solution
33