6. Enter any Required or Optional information while
the analysis is running.
NOTE: The sample will remain on screen until all Required fields
have been entered.
The analysis can
be cancelled by
pressing the X icon
will not be printed
or transmitted.

3.1.2 Utilizing the Luer Station

The luer station provides a means of attaching a syringe
to the analyzer instead of manually positioning the sample
probe in the sample. When utilizing the leur station Nova
recommends using the Nova Syringe Clot Catcher to ensure
that the sample is positioned correctly for aspiration and to
prevent clots from entering the flowpath. If a clot catcher
is not used, syringes must be filled with sufficient sample
for the probe to travel approximately 1-inch (26 mm) into
the syringe.
From the Home screen:
From the Home screen, login if you are prompted to login.
1. Press the Login icon
2. Enter or scan your Operator ID then press
3. If required, enter or scan your Password then press
but results
Figure 3.5 Sample Information Screen
.
3 Sample Analysis
to log into the analyzer.
.
3-3