Prime CCS Instructions for Use Manual
Sample
QC Level 1
QC Level 2
QC Level 3
Linearity Std 1
Linearity Std 2
Linearity Std 3
Linearity Std 4
Linearity Std 5
A.3 Calibrator Cartridge
The concentrations of the internal standards are printed on
the Calibrator Cartridge. In addition to the calibrators and
solutions, the Calibrator Cartridge has a self-contained
waste bag for safe disposal of waste.
A.3.1 Traceability of Calibrators, Controls, and Standards
Chemistry analytes are traceable to the Standard Reference
Materials of the National Institute of Standards and
Technology (NIST). SO
A-22
Lactate Precision Data
Pooled
Within run
Mean
N
SD (Sr)
(mmol/L)
0.79
240
0.022
2.60
240
0.012
6.38
240
0.023
13.36
240
0.229
0.67
240
0.013
2.53
240
0.022
6.59
240
0.056
10.51
240
0.086
is traceable to tonometry.
2
Total
Within
imprecision
run % CV
SD (St)
2.77
0.03
0.47
0.01
0.35
0.06
1.71
0.79
1.97
0.08
0.87
0.07
0.85
0.26
0.82
0.45
Total
Imprecision
%CV
3.58
0.54
0.87
5.91
12.71
2.88
3.92
4.26