5.2 Reference Sensor Replacement

1. From the Home screen, press the Tool Box
icon.
2. From the System Menu select the Change Reference
Sensor and wait for the pump to stop.
3. Open the door; open Cartridge door.
4
Remove Sensor Card.
NOTE: To change the Reference Sensor the Sensor Card
must be removed first. Only touch Sensor Card by
the edges.
Pump Tubing
Manifold
Sensor Card
Reference
Sensor
Reference
Sensor Arm
S-Line
Figure 5.8 Reference Sensor
5 Periodic Replacements
5-5