Welch Allyn CP 100 Update To Directions For Use page 38

Electrocardiographs
Hide thumbs Also See for CP 100:
Table of Contents

Advertisement

48
Chapter 5 Performing ECG Tests
Recording a Stat Auto ECG
A stat Auto ECG is an immediate printout in Auto Report 1 format.
Stat mode bypasses patient data entry, as shown in
number is assigned to the patient to identify stat tests. After printing, you may enter the
patient data by editing the test.
In stat mode, Auto Send and Auto Save are always disabled, even if they are enabled in
the ECG settings. If you want to send or save a stat Auto ECG, you may do so manually
after you print.
To Record a Stat Auto ECG
1. Press and hold the Auto ECG key
The electrocardiograph begins acquiring ECG data. After it has acquired 10 seconds of
quality data, it prints a report.
Note
2. Go to
Continue the procedure as if you have just pressed the Print ECG softkey.
WARNING To avoid the risk of associating ECG reports with the wrong patients,
each report must identify the patient. If any ECG report does not identify the
patient, write the patient identification information on the report immediately
following the ECG test.
Write the patient identification information on the printed report immediately after
the ECG test to avoid associating the ECG report with the wrong patient.
ECG report Print Copy feature will not be available if no Patient Name or ID are
entered.
Step 7
on page 58.
Welch Allyn CP 200 Electrocardiograph
Figure 36
on page 56. A temporary ID
.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cp 200

Table of Contents