9. Press Exit to complete patient testing.
To Edit or Confirm a Test Directly After Printing
Note
1. From the Auto ECG Post-Print screen
The Edit Test – Patient Data screen appears.
2. (Optional) Edit the patient data.
3. Press the desired softkey.
Softkey
Interp
Cancel
Done
WARNING Failure to follow these instructions could incorrectly associate an
ECG test with the wrong patient.
Caution The requirements of AAMI EC11, Section 3.2.7.2, Frequency and
Impulse Response, for an impulse triangle waveform may be impacted by up to 5
milliseconds of small amplitude dampened ringing immediately after the impulse
when the muscle filter (35 Hz) is turned on or a small amplitude offset when the
baseline filter (0.5 Hz) is turned on. These requirements are unaffected by any
other combination of filters turned on or off. Measurements performed by the
optional interpretation algorithm are unaffected by any filter selections.
A qualified physician must review and confirm all tests before patients are
treated. If changes are needed for any saved test, you can edit two types of data:
patient data
interpretation statements
Effect
Saves any changes, and displays the Edit Test – Interpretation screen.
1. (Optional) Edit the interpretation statements to be saved with the test.
2. Press the desired softkey:
• Patient Data saves any changes, and returns to the Edit Test – Patient Data
screen.
• Confirm saves any changes, sets the test status to "confirmed," and returns to the
Auto ECG Post-Print screen.
• Cancel discards any changes, and returns to the Auto ECG Post-Print screen.
• Done saves any changes, and returns to the Auto ECG Post-Print screen.
For details on interpretation, see
Discards any changes, and returns to the Auto ECG Post-Print screen.
Saves any changes, and returns to the Auto ECG Post-Print screen.
Chapter 5 Performing ECG Tests
(Figure 39
on page 59), press Edit Test.
"Automatic ECG interpretation"
47
on page 5.