9.3

Troubleshooting

9.3.1
No power to the darkroom or transilluminator
1.
Recheck the main power cord connection to the darkroom as well as the power cables
between the darkroom and transilluminator, LED White Light Plate or optional UV
handheld lamps.
2.
Check the fuses located at the back of the unit next to the power port. A small flat-head
screwdriver or similar tool will be required. Turn the fuseholder cap counterclockwise
and the fuse holder will pop out. Inspect the thin wire within the glass fuse to see if there
is a break in the wire. If so, replace the fuse(s). If fuses are blowing repeatedly, contact
UVP Technical Support for additional troubleshooting.
9.3.2
A partial image appears on the screen
1.
Ensure that the filter wheel is not off center by turning the filter wheel selector until it
fully engages or "clicks" into place.
2.
Remove the camera/lens combination and verify that the emission filter is fully seated in
the filter wheel.
9.3.3
Transilluminator will not turn on
1.
In addition to the Ultraviolet: Trans power switch on the front of the darkroom, the
transilluminator itself has a power switch. Make sure that the green transilluminator
power switch, located on the front of the transilluminator, is in the ON position.
2.
For UV protection and to extend the life of the UV transilluminator, there is a ten-minute
transilluminator shutoff timer installed in the system. The timer shuts off the
transilluminator after ten minutes of operation. Push the reset button on the front of the
darkroom to reset the timer.
9.3.4
Error messages appear on the screen
1.
An error message that is related to the VisionWorks software interface or Microsoft
Windows may appear on the screen. If the message is related to Microsoft Windows,
such as a reminder to activate or update the copy of Windows, please contact your sys-
tem administrator for assistance.
2.
If an error message appears repeatedly and your system administrator does not
recognize it as a Microsoft Windows error, contact UVP Technical Support for further
assistance.
10

Disposal

At the end of its service life the Imaging System and all its electronic components must be
disposed of in accordance with the applicable regulations as electronic waste.