Calibration - Rice Lake 520 Installation Manual

Digital weight indicator
Table of Contents

Advertisement

4.0

Calibration

The 520 can be calibrated using the front panel, EDP commands, or the Revolution
method consists of the following steps:
Zero calibration
Entering the test weight value
Span calibration
Optional five-point linearization
Optional rezero calibration for test weights using hooks or chains.
The following sections describe the calibration procedure for each of the calibration methods (see Section 4.1 on
page 37 for further explanation on calibration).
CONFIG
XXXXXXX
FORMAT
LATITUD
45
0–90
If GRAVADJ = ON (PROGRM Menu)
WLIN
PT–> 1
Display and edit
test weight value
Enter mV
If MVCALIB = ON
(PROGRM Menu)
CAL*
*
Display and edit
linear point
calibration
A/D count value
36
520 Installation Manual
CALIBR
SERIAL
ELEVATN
WZERO
Enter mV
+0345
+
-
9999 - 9999
If MVCALIB = ON
(PROGRM Menu)
CAL*
*
Display and edit
zero calibration
A/D count
value
PT–> 2
PT–> 3
Same as PT-> 1
Figure 4-1. Calibration (CALIBR) Menu
PROGRM
XXXXXXX
PFORMT
DIG IN
WVAL
Display and edit
test weight value
PT–> 4
PT–> 5
configuration utility. Each
XXXXXXX
ALGOUT
XXXXXXX
VERS
WSPAN
Enter mV
If MVCALIB = ON
(PROGRM Menu)
CAL*
*
Display and edit
span calibration
A/D count
value
REZERO
CAL*
*
Press Enter to
remove offset from
zero and span
calibrations