Using Labware MiniHub Diagnostics in the VWorks software
Rotating the MiniHub to a specific position
10
Speed
Description
Slow
Rotates the MiniHub at 25% of the factory- set
maximum speed.
Medium
Rotates the MiniHub at 50% of the factory- set
maximum speed.
Fast
Rotates the MiniHub at 100% of the factory- set
maximum speed.
You can rotate the Labware MiniHub using one of the following methods:
Rotating to a specific cassette
Rotating the MiniHub in set increments
Rotating the MiniHub to a specific position relative to home
CAUTION
Obstructing the Labware MiniHub while it is rotating will cause an error.
Remove any obstacle that might prevent the Labware MiniHub from rotating.
Rotating to a specific cassette
You can rotate the MiniHub so that the desired cassette is facing the robot.
To rotate to a specific cassette:
1
In the Controls tab, select the desired cassette from the Rotate to Cassette list,
and then click Rotate to Cassette. The MiniHub rotates until the desired
cassette faces the robot. The Cassette box displays the selected cassette.