Contact 4 Against Vessel; Measured Value Invalid - Vega EL 11 Operating Instructions Manual

Capacitive electrodes el series 4...20 ma - compact
Table of Contents

Advertisement

Diagnosis
VVO-
failure message
Meas. value invalid
Capacitive electrodes EL 4 ... 20 mA - compact

Contact 4 against vessel

The connection between contact 4 and the metal vessel (not instrument
hexagon or electrode flange) should be as good as possible. Measure with
an ohmmeter (range very low) the resistance value between contact 4 and
the vessel.
• Shortcircuit (0 ... 3 Ω), optimum connection
• Resistance > 3 Ω
- corrosion on the mounting boss or flange
- probably the mounting boss was covered with Teflon tape or similar
Check the connection to the vessel. If there is no connection, you can
connect a line from the earth terminal outside to the vessel.
Note that coated flanges must be in any case connected via the earth
terminal to the vessel.
Contact 4 against 6
In case of values > 3 Ω there is a defect.
When you cannot find a failure in the electrode, than exchange the
oscillator against a similar replacement type (if available) or send the
electrode for repair to VEGA.
After insertion of the new electronics, carry out an adjustment. See 4 Set-
up.

Measured value invalid

The actual measured value is not within the valid measuring range.
Reasons can be e.g. extreme modifications of the dielectric constant figure
or wrong adjustments.
Modify the actual measuring conditions.