Repairing the Column Compartment
7 Fix the Z-panel with the two screws.
9 Carefully replace the heat exchanger cables into the foam
channels.
86
8 Reconnect the heat exchanger cables to the processor
board and the grounding connection of the column
switching valve.
Next steps:
10 Replace the foam section, the top cover and front
cover, see
"Installing the Foam and the Top Cover"
on page 96.
11 Replace the column compartment into stack.
12 Reconnect capillaries.
13 Reconnect the power cable.
14 Turn on the column compartment.