Shutdown

An external instrument has generated a shut-down signal on the remote line.
The instrument continually monitors the remote input connectors for status
signals. A LOW signal input on pin 4 of the remote connector generates the
error message.
Probable Causes
• Leak detected in an external instrument with a remote connection to the
• Shut-down in an external instrument with a remote connection to the
• The degasser failed to generate sufficient vacuum for solvent degassing.
Suggested Actions
1100 Series TCC Reference Manual
system.
system.
Fix the leak in the external instrument before restarting the pump module.
Check external instruments for a shut-down condition.
Check the degasser module for an error condition. Refer to the Reference
Manual for the Agilent 1100 Series vacuum degasser.