11 Position the leak sensor so that it does not touch the
bottom of the leak base.
13 Replace the leak tubing assembly into its location on leak
base. Ensure that it keeps in this position during the next
steps.
1100 Series TCC Reference Manual
Repairing the Column Compartment
12 Replace the clips (shown from the inside of the column
compartment) on both sides using a flat screwdriver.
14 Replace the heat exchanger assemblies, see
the Heat Exchanger Assemblies"
4
"Installing
on page 84.
93