Agilent Technologies 1100 Series Reference Manual page 190

Thermostatted column compartment
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

7
Control Module Screens for the Agilent 1100 Column Compartment
Use the Esc key to receive Views on the F5 key. Choose System from the
pull-down menu. Use the F3 key (Records) to select the Generic module. In
this screen the resident firmware revision is shown.
Use the F5 key (FW Update) to enter the Update section. Select the a file from
the PCMCIA card (1316nnnn.DLB) and press execute. When the update has
finished, press F7 key (Transfer) to return the module into the normal mode
(LED on module should stay yellow).
If you have not saved your methods, please do it before continuing. Otherwise
they will be overwritten during the update process.
180
1100 Series TCC Reference Manual