Gpib Interface; Can Interface; Remote Interface - Agilent Technologies 1100 Series Reference Manual

Thermostatted column compartment
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

6
Introduction to the Column Compartment

GPIB Interface

The GPIB connector is used to connect the module with a computer. The
address and control switches next to the GPIB connector determine the GPIB
address of your module. The switches are preset to a default address and
recognized by the operating software from Agilent Technologies.
Table 34
Autosampler
Pump
FLD
VWD
Agilent 8453A
DAD/MWD
Column Compartment

CAN Interface

The CAN is an intermodule communication interface. It is a 2 wire serial bus
system supporting high speed data communication and real-time requirement.

Remote Interface

The APG Remote connector may be used in combination with other analytical
instruments from Agilent Technologies if you want to use features as common
shut down, prepare, and so on.
Remote control allows easy connection between single instruments or systems
to ensure coordinated analysis with simple coupling requirements.
The subminiature D connector is used. The module provides one remote
connector which is inputs/outputs (wired-or technique).
To provide maximum safety within a distributed analysis system, one line is
dedicated to SHUT DOWN the system's critical parts in case any module
detects a serious problem. To detect whether all participating modules are
switched on or properly powered, one line is defined to summarize the
POWER ON state of all connected modules. Control of analysis is maintained
154
Default Addresses
28
22
23
24
25
26
27
Autosampler
RID
Autosampler (HP 1050)
Pump (HP 1050)
VWD (HP 1050)
DAD (HP 1050)
1100 Series TCC Reference Manual
28
29
11
12
10
17

Hide quick links:

Advertisement

Table of Contents
loading